File Exchange

image thumbnail

RAPID: a Routine Assurance Pipeline for Imaging of Diffusion

version 1.2.0.0 (27.5 KB) by Silvia
uses diffusion MRI data for Quality Assurance of diffusion acquisitions

0 Downloads

Updated 05 Jul 2012

View License

RAPID: a Routine Assurance Pipeline for Imaging of Diffusion - De Santis et al. (submitted)

script_to_calculate_SNR calculates the signal-to-noise ratio and returns the optimal parameters for acquiring QA data.
_Input: nifti files of 100 b=0 images
_Output: max b-value and voxel size

script_to_run_QA checks for the linearity of b, the uniformity of Gmax across the field-of-view, the mutual agreement of gradient power across the three logical axes and corrects for gradient mismatches.
_Input: nifti files of diffusion data on phantom acquired using the gradient table Grad_dirs_QA_shuffled.txt
_Output: .mat file of QA with date

script_to_compare_QA_results checks for temporal stability.
_Input: two .mat files of QA results

Comments and Ratings (0)

Updates

1.2.0.0

4 Jul 2012 - correction for imaging gradients, to be used with the same geometry of the actual scan to be corrected

1.1.0.0

UPDATE: included a tool to account for imaging gradients in the b-matrix calculation